Jalali M T, Shahbazian H B, Afsharmanesh M R, Mousavi Dehmordi R, saki A. Evaluation of Accuracy, Precision and Agreement of Five Hba1c Measurement Methods with HPLC Reference Method. mljgoums 2016; 10 (2) :58-64
URL: http://mlj.goums.ac.ir/article-1-852-en.html
1- Department of Laboratory Sciences, School of Paramedical Sciences, Jundishapur University of Medical Sciences, Ahvaz, Iran
2- Health Research Institute, Diabetes Research Center, Jundishapur University of Medical sciences, Ahvaz, Iran
3- Faculty of Medicine, Jundishapur University of Medical sciences, Ahvaz, Iran , mra448@yahoo.com
4- Faculty of Medicine, Jundishapur University of Medical Sciences, Ahvaz, Iran
5- Jundishapur University of Medical Sciences, Ahvaz, Iran
Abstract:   (13788 Views)

ABSTRACT

        Background and Objective: The current challenge of diabetes mellitus is to prevent its complications. These complications are directly associated with hyperglycemia in diabetics. The HbA1c measurement is essential for long-term glycemic control. Synchronization of HbA1c measurement is important in order to avoid discrepancies between results reported by laboratories. This study aimed to evaluate the accuracy, precision and agreement of five HbA1c measurement methods with HPLC reference method.

       Methods: HbA1c levels of 55 samples were measured using six methods of microcapillary electrophoresis (Sepia), enzymatic method (Pishtaz Teb), immunoturbidometry (Pars Azmoon), boronate affinity (Nycocard), immunofluorescence (ichroma) and Tosoh G8 HPLC.

       Results: The five tested methods showed a good agreement with the HPLC method with correlation coefficient of less than 95%. Regression testing of HPLC method and other methods showed slope of 0.99 (P<0.05) for Sebia, 1.02 (P<0.05) for Pishtaz Teb, 0.79 (P<0.05) for Pars Azmoon, 0.82 (P<0.05) for Nycocard and 0.89 (P<0.05) for ichroma. Average inaccuracy for the Sebia, Pishtaz Teb, Pars Azmoon, Nycocard and ichroma in comparison with the HPLC reference method were -0.09, -0.004, -0.75, -0.79 and -0.78, respectively.

         Conclusion: The Sebia microcapillary method and Pishtaz teb enzymatic method have appropriate accuracy and precision. Therefore, these methods can be used as alternatives to the HPLC method for HbA1c measurement. Other methods such as Pars Azmoon, Nycocard and ichroma have significant shortcomings in terms of accuracy.

     

Full-Text [PDF 529 kb]   (3470 Downloads)    
Research Article: Original Paper |
Received: 2016/07/25 | Accepted: 2016/07/25 | Published: 2016/07/25 | ePublished: 2016/07/25

Add your comments about this article : Your username or Email:
CAPTCHA

Send email to the article author


Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.

© 2007 All Rights Reserved | Medical Laboratory Journal

Creative Commons License
This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.